Article details

Technical parameters Of IEC60068-2-75 spring hammer calibration appartus

日期:2024-05-05 02:04
浏览次数:536
摘要: Technical parameters Of IEC60068-2-75 spring hammer calibration appartus Principle of calibration: The principle of this calibration procedure is to compare the energy provided by a spring hammer, which is difficult to measure directly, to the energy of a pendulum, calculated from its mass and height of fall. And it's designed with the requirements of Annex B of IEC60068-2-75. Construction: It's maily made of a frame, a bearing, a drag

Technical parameters Of IEC60068-2-75 spring hammer calibration appartus

Principle of calibration:

The principle of this calibration procedure is to compare the energy provided by a spring hammer, which is difficult to measure directly, to the energy of a pendulum, calculated from its mass and height of fall. And it's designed with the requirements of Annex B of IEC60068-2-75.

 

Construction:

It's maily made of a frame, a bearing, a drag pointer, a pendulum, a release base and a release device.

 

Notes:

The Calibration Device fo Spring Hammer can be used for all models, allowing verification of the proper functioning of different types of hammers with the same calibration.

 The spring is designed for calibrating spring hammers having characteristics as defined for energy values equal to less than 2J.

 

Conforms to:

IEC60335, IEC60065, IEC60745, IEC61029, IEC60601, IEC60610 etc.

 

Technical parameters:

The actual accuracy: 0.01J

Repeat accuracy: ±0.01J

Maximum range: 0 to 2J

Guide groove diameter: 51mm

Pendulum energy loss: less than 0.002j

Trigger route: more than 30mm

Weight: 1pcs