a brief introdction:
IEC 60112 CTI PTI Tracking Flammability Test Chamber The leakage tracking test equipment is designed and manufactured according to the standard requirements of IEC60112, it is required by IEC60335-1 clause 29.2 and Annex N, IEC 60065:2014 clause 13.4, IEC60598-1, IEC 60884-1, UL746A and ASTM D3638-92 and etc. It is used for determination of the relative resistance to tracking performance that so
Details description:


IEC 60112 CTI PTI Tracking Flammability Test Chamber 


Product overview:

The leakage tracking test equipment is designed and manufactured according to the standard requirements of IEC60112, it is required by IEC 60065:2014 clause 13.4, IEC60598-1, IEC 60884-1, UL746A and ASTM D3638-92 and etc. It is used for determination of the relative resistance to tracking performance that solid electrical insulating material under the action of an electric field, when the surface exposed to water containing impurities.

 

The leakage tracking tester adopts to advanced PLC control system and humanized 7” color touch screen, the dripping is controlled by a high-precision micro water supply pump, the timing is automatic, the electrode pressure sensor is accurate, and the operation is simple, accuracy is high, it is leading among similar equipment.

 

Technical parameters:

Parameter model CX-LD19A
Working voltage Match client's local power supply requirements
Control Advanced PLC
Display 7 inch color touch screen
Testing voltage 0~600V adjustable, precision 1.5%
Timing device 9999X0.1S
Electrode Platinum electrodes and brass electrodes, each of a pair
Electrode dimension

(5±0.1)×(2±0.1)×(≥12),

30°slant

Electrode included angle and distance 60°±5°, distance is (4±0.1) mm
Electrode pressure 1.00N±0.05N
The interval of dropping liquid 30±5S, adjustable
Dropping liquid height 35±5mm
Dropping liquid delay 0.2S(solenoid valve opening time)
Dropping liquid volume 45~50 drip/cm3
Time-delay circuit 2±0.1S (in 0.5A or larger current)
Short-circuit pressure drop 8% MAX
Short-circuit current

1±0.1A

Accuracy: 1%

Wind speed 0.2m/s
Environmental requirement 0~40℃, relative humidity≤80%, in the place of no obvious vibration and corrosive gas
Equipment size & weight 1100mm* 620mm* 1450mm, 180KGS
conforms to standard IEC60695, IEC60112, IEC 60598-1, IEC 60884-1, UL746A, ASTMD3638, etc.
IEC 60112 CTI PTI Tracking Flammability Test Chamber PLC Control 0